Carbon Dioxide Detector

Editor: Shawn London Updated: 1/29/2023 9:21:39 AM

Introduction

The measurement of carbon dioxide (CO2) was first developed in the early 1900s; however, it was complex and of limited clinical use.[1] On the most basic level, end-tidal CO2 (ETCO2) detectors measure the exhaled concentration of carbon dioxide. The concentration of exhaled COcorrelates to the arterial concentration of CO2, although multiple factors can impact the relationship. Technological advances have allowed CO2 to become a common measurement in healthcare with many clinical applications.

Although CO2 detectors ultimately measure exhaled CO2, some units have design differences. Detectors can be divided into quantitative and qualitative types. Qualitative detectors demonstrate the presence or absence of CO2. The most basic unit of this type is the colorimetric detector, producing a color change as the gas flows through a litmus paper film in the airway circuit. Qualitative detectors are primarily used to initially confirm the proper placement of an endotracheal tube by demonstrating the presence (connoting successful placement in the trachea) or lack (esophageal placement) of exhaled CO2. These detectors are usually elementary, light, and portable devices that do not require power. Quantitative detectors measure the level of CO2 that is exhaled and, therefore, provide more detailed information. Qualitative detection can be further subdivided into capnography and capnometry. Capnography produces a waveform in a graphic format, while capnometry provides a numeric display of the CO2 concentration. Waveform capnography provides more information to the clinician. Alterations of the waveform morphology can be interpreted to give clues to the clinical situation.

Their location in the airway circuit is also used to categorize detectors. These are typically divided into mainstream and sidestream devices. Mainstream detectors utilize an IR sensor positioned directly within the gas flow path and produce real-time graphs of the CO2 waveform. These tend to be more costly units used primarily for intubated patients. Sidestream detectors are located off the main path of flow in the airway circuit. Sidestream detectors often use small-diameter tubing leading to a detector that allows gas sampling. Nasal monitoring of end-tidal CO2 typically makes use of this configuration.

There are some pitfalls and considerations when using CO2 detectors. Fluids such as vomitus, frothy secretions from a patient experiencing acute pulmonary edema, sputum from pneumonia, or moisture can render CO2 detectors ineffective. The detector paper in colorimetric units is particularly sensitive to fluids. Mainstream detectors are located directly in the path of the airway circuit and, therefore, increase the volume of undesirable dead space. On the other hand, Sidestream detectors may experience a delay compared to mainstream detectors because of the remote location of the sensor.

Anatomy and Physiology

Register For Free And Read The Full Article
Get the answers you need instantly with the StatPearls Clinical Decision Support tool. StatPearls spent the last decade developing the largest and most updated Point-of Care resource ever developed. Earn CME/CE by searching and reading articles.
  • Dropdown arrow Search engine and full access to all medical articles
  • Dropdown arrow 10 free questions in your specialty
  • Dropdown arrow Free CME/CE Activities
  • Dropdown arrow Free daily question in your email
  • Dropdown arrow Save favorite articles to your dashboard
  • Dropdown arrow Emails offering discounts

Learn more about a Subscription to StatPearls Point-of-Care

Anatomy and Physiology

Waveforms

The graph demonstrates a typical CO2 waveform that might be seen from a mainstream capnograph (see Graph. CO2 Waveform). Phase A represents the first exhalation step as gas from the dead space arrives at the detector. This dead space gas contains no CO2 introduced from the lungs and, therefore, is seen as a flat, zero concentration of CO2 over time. Phase B demonstrates the rapid rise in CO2 as the expiratory phase continues and the concentration of CO2 rises. This phase reflects a mixing of the dead space gas, which lacks CO2, and the alveolar gas. Phase C represents the alveolar plateau phase, where the concentration of CO2 becomes more constant over time. The rate of rise of this phase can change in certain pathological states, for example, with bronchospasm, where the rise is steeper. Point D represents the “end-tidal CO2” level; this is the highest concentration of CO2 during the cycle and occurs at the end of expiration. Phase E shows the rapid decline in CO2 concentration that begins with inspiration as external gas enters the lungs. This external gas has a low concentration of CO2, and the waveform returns to baseline.

Alterations to this typical waveform can be used as clues that alert the clinician to changes in the status of their patient. Hyperventilation typically produces a pattern of increased respiratory frequency associated with a depressed waveform. Hypoventilation is known to occur in at least 2 common types. A decreased respiratory rate causes Bradypnea hypoventilation and produces a waveform pattern with low respiratory frequency and elevated ETCO2. Hypopneic hypoventilation, in contrast, reveals a decreased ETCO2 caused by the patient taking shallower breaths than normal. Although these are both forms of hypoventilation, 1 results in an elevation of ETCO2, and the other causes a decrease in ETCO2.

Clinical Significance

Procedural Sedation and Analgesia

Continuous end-tidal CO2 monitoring during the performance of procedural sedation and analgesia (PSA), sometimes called “conscious sedation,” is often recommended as a measure to detect respiratory depression. The medications associated with PSA, such as proposal, benzodiazepines, and narcotics, can often lead to respiratory depression. It is essential to recognize that the levels of sedation represent a continuum of drug effects. The lightest forms of sedation include mild anxiolysis, where a patient can generally respond after receiving these medications. At the other extreme is deep sedation or general anesthesia, where the patient requires active respiratory support and may require hemodynamic support during the procedure. PSA is intended to preserve a patient’s basic cardiopulmonary functioning; however, the medications used in PSA can have side effects, including hypotension, loss of protective airway reflexes, and hypoventilation. Respiratory depression is considered to be 1 of the most significant risks during the administration of PSA. To address the risk of hypoventilation, many experts who study and provide PSA have explored whether CO2 monitoring can help identify hypoventilation and improve patient safety. Oxygen saturation has classically been used as the marker for hypoventilation. However, this may be a late marker of hypoventilation, particularly when supplemental oxygen is administered prophylactically during the administration of PSA.

Some specialty societies whose members perform PSA have recommendations about CO2 monitoring.[2] The American Society for Anesthesia promotes stronger recommendations about CO2 monitoring during moderate and deep sedation while others have less proscriptive suggestions about its use.[3][4][3] Adoption of the practice of regular CO2 monitoring during PSA has been less rapid for those providers outside of the operating room due to limitations in evidence to support objective benefits and the lack of available equipment and training.[5] Furthermore, the clinical relevance is still being evaluated.[6] Other groups, such as a recent Cochrane review, have argued that the current evidence is not able to demonstrate a difference in outcomes when using CO2 monitoring in addition to standard monitoring.[7]

Endotracheal Tube Confirmation and Monitoring

CO2 detection has grown to be used as an important indicator to confirm the correct placement of an endotracheal tube (ETT) after intubation. Airway experts widely recommend this as a required step in ETT confirmation.[8] Capnography has been shown to be a highly reliable method of endotracheal tube confirmation. A properly positioned tube in the trachea produces a reliable CO2 waveform that can be detected reliably by these sensors. Esophageal intubation does not transmit CO2 to the detector and would indicate improper placement. Hypopharyngeal intubation is a special circumstance when the endotracheal tube is not positioned through the vocal cords, but rather just superior to the glottis. Hypopharyngeal intubations can be difficult to identify clinically, and capnography may assist in detecting this situation through an abnormal or inconsistent waveform. Additional methods of confirming tube position include direct visualization of the glottis, auscultation for breath sounds and inspection for chest rise during ventilation.

A false negative CO2 reading may occur in cases of cardiac arrest when there is no cardiopulmonary circulation to carry CO2 to the alveoli and, ultimately, to the detector. In this case, it is possible that an endotracheal tube could be properly positioned in the trachea yet not provide a CO2 waveform. However, the detector does not produce a waveform with ventilation and could lead to the wrong conclusion. Early initiation of CPR during cardiac arrest leads to circulatory flow which can then generate aO22 waveform, helping confirm proper tube position. In the later phases of resuscitation as cell death occurs, even with CPR being performed, the CO2 levels fall.

Continuous CO2 monitoring of an intubated patient can also help identify accidental endotracheal tube displaced during transport or movement of a patient. Continuous monitoring can also help detect other post-intubation complications such as airway obstruction from a kinked ETT or mucous plugging. Anytime a waveform is suddenly lost, it should alert the clinician to evaluate the equipment and connections and determine if there is a more concerning cause for the loss.

Cardio Pulmonary Resuscitation (CPR)

Effectiveness of CPR

Immediately after a cardiac arrest and during CPR, chest compressions generate cardiac output that results in circulatory flow and the generation of a CO2 waveform. In cardiac arrest without compressions, there may be no detectable CO2 level. End-tidal CO2 levels correlate with the effectiveness of the compressions and can be used to help guide resuscitation.[9] High-quality CPR produces higher cardiac output and, therefore, higher CO2 levels than poor-quality CPR.  This information can guide the resuscitation. Important factors that lead to high-quality CPR include rate, depth, recoil and the lack of interruptions. These factors should be assessed when there is a low CO2 level during the resuscitation of a patient with cardiac arrest. Compressor fatigue can lead to loss of quality compressions and can be detected by a decline in end-tidal CO2 levels. Cardiac arrest teams can give feedback to the compressor about their technique, or they may choose alternate to a “fresh” compressors based on these levels. The American Heart Association 2010 recommendations suggest that an ETCO2 level of less than 10 mm Hg during CPR would be a reasonable indication to aim efforts toward improving the quality of compressions.

Return of Spontaneous Circulation

During cardiac arrest, when there is a triumphant return of spontaneous circulation (ROSC), there is an associated rise in end-tidal CO2 levels; this is often the first sign of ROSC.[10] A sudden rise in CO2 may alert the cardiac arrest team that there is ROSC and trigger an assessment of circulatory status through pulse, rhythm, and blood pressure. The clinical relevance is to avoid unnecessary interruptions in CPR. A rise in CO2 may be an earlier sign of ROSC than performing frequent pulse checks that cause a pause in compressions.

Termination of Resuscitation

End-tidal CO2 is also helpful during resuscitation to help predict death after a prolonged cardiac arrest. After 20 minutes of CPR, an end-tidal CO2 level of 19 mm Hg or less is predictive of death as an outcome of the cardiac arrest.[11] It can be challenging to make the clinical decision when to terminate resuscitative efforts when caring for a patient experiencing cardiac arrest. ETCO2 levels provide additional data to the clinician to help make this decision.[12]

Future and Growing Applications

ETCO2 approximates the arterial concentration of CO2 (PaCO2).  It is typically 2 to 5 mm Hg lower and can be used to estimate the effectiveness of ventilation. This gap can be affected by multiple disease states such as asthma, COPD, pneumonia, pulmonary embolism among others that cause a change in V/Q mismatching, alteration of physiological dead space or shunt physiology. CO2 detection provides a window into ventilation, pulmonary perfusion, and cardiac output.  Multiple disease states are being studied in attempts to find clinical correlations to CO2 levels. Recent work has explored various possible uses, including determination of metabolic rate, diagnosis of pulmonary embolism, and assessment for diabetic ketoacidosis.[13][14][15] As equipment becomes more available, novel uses of capnography are being studied and the relationship between end-tidal carbon dioxide and CPR quality during both in-hospital and out-of-hospital cardiac arrest.

Enhancing Healthcare Team Outcomes

Nurses, respiratory therapists, and clinicians must be aware of the pitfalls and considerations when using CO2 detectors. an interprofessional team approach to monitoring patients using these devices produces the best patient outcomes.

Media


(Click Image to Enlarge)
<p>CO2 Waveform</p>

CO2 Waveform

Created and Contributed by T Nowicki, MD

References


[1]

Siobal MS. Monitoring Exhaled Carbon Dioxide. Respiratory care. 2016 Oct:61(10):1397-416. doi: 10.4187/respcare.04919. Epub 2016 Sep 6     [PubMed PMID: 27601718]


[2]

Godwin SA, Burton JH, Gerardo CJ, Hatten BW, Mace SE, Silvers SM, Fesmire FM, American College of Emergency Physicians. Correction: Correction to 'Clinical Policy: Procedural Sedation and Analgesia in the Emergency Department' [Annals of Emergency Medicine 63 (2014) 247-258.e18]. Annals of emergency medicine. 2017 Nov:70(5):758. doi: 10.1016/j.annemergmed.2017.03.042. Epub 2017 Apr 7     [PubMed PMID: 28395927]


[3]

. Practice guidelines for sedation and analgesia by non-anesthesiologists. A report by the American Society of Anesthesiologists Task Force on Sedation and Analgesia by Non-Anesthesiologists. Anesthesiology. 1996 Feb:84(2):459-71     [PubMed PMID: 8602682]

Level 1 (high-level) evidence

[4]

American Society of Anesthesiologists Task Force on Sedation and Analgesia by Non-Anesthesiologists. Practice guidelines for sedation and analgesia by non-anesthesiologists. Anesthesiology. 2002 Apr:96(4):1004-17     [PubMed PMID: 11964611]

Level 1 (high-level) evidence

[5]

Conway A, Douglas C, Sutherland JR. A systematic review of capnography for sedation. Anaesthesia. 2016 Apr:71(4):450-4. doi: 10.1111/anae.13378. Epub 2016 Jan 21     [PubMed PMID: 26792775]

Level 1 (high-level) evidence

[6]

Smally AJ, Nowicki TA, Simelton BH. Procedural sedation and analgesia in the emergency department. Current opinion in critical care. 2011 Aug:17(4):317-22. doi: 10.1097/MCC.0b013e328348bf43. Epub     [PubMed PMID: 21677578]

Level 3 (low-level) evidence

[7]

Wall BF, Magee K, Campbell SG, Zed PJ. Capnography versus standard monitoring for emergency department procedural sedation and analgesia. The Cochrane database of systematic reviews. 2017 Mar 23:3(3):CD010698. doi: 10.1002/14651858.CD010698.pub2. Epub 2017 Mar 23     [PubMed PMID: 28334427]

Level 1 (high-level) evidence

[8]

Knapp S, Kofler J, Stoiser B, Thalhammer F, Burgmann H, Posch M, Hofbauer R, Stanzel M, Frass M. The assessment of four different methods to verify tracheal tube placement in the critical care setting. Anesthesia and analgesia. 1999 Apr:88(4):766-70     [PubMed PMID: 10195521]

Level 1 (high-level) evidence

[9]

Sheak KR, Wiebe DJ, Leary M, Babaeizadeh S, Yuen TC, Zive D, Owens PC, Edelson DP, Daya MR, Idris AH, Abella BS. Quantitative relationship between end-tidal carbon dioxide and CPR quality during both in-hospital and out-of-hospital cardiac arrest. Resuscitation. 2015 Apr:89():149-54. doi: 10.1016/j.resuscitation.2015.01.026. Epub 2015 Jan 30     [PubMed PMID: 25643651]

Level 2 (mid-level) evidence

[10]

Pokorná M, Necas E, Kratochvíl J, Skripský R, Andrlík M, Franek O. A sudden increase in partial pressure end-tidal carbon dioxide (P(ET)CO(2)) at the moment of return of spontaneous circulation. The Journal of emergency medicine. 2010 Jun:38(5):614-21. doi: 10.1016/j.jemermed.2009.04.064. Epub 2009 Jul 1     [PubMed PMID: 19570645]

Level 2 (mid-level) evidence

[11]

Levine RL, Wayne MA, Miller CC. End-tidal carbon dioxide and outcome of out-of-hospital cardiac arrest. The New England journal of medicine. 1997 Jul 31:337(5):301-6     [PubMed PMID: 9233867]


[12]

Poon KM, Lui CT, Tsui KL. Prognostication of out-of-hospital cardiac arrest patients by 3-min end-tidal capnometry level in emergency department. Resuscitation. 2016 May:102():80-4. doi: 10.1016/j.resuscitation.2016.02.021. Epub 2016 Mar 3     [PubMed PMID: 26948821]


[13]

Bou Chebl R, Madden B, Belsky J, Harmouche E, Yessayan L. Diagnostic value of end tidal capnography in patients with hyperglycemia in the emergency department. BMC emergency medicine. 2016 Jan 29:16():7. doi: 10.1186/s12873-016-0072-7. Epub 2016 Jan 29     [PubMed PMID: 26821648]


[14]

Manara A, D'hoore W, Thys F. Capnography as a diagnostic tool for pulmonary embolism: a meta-analysis. Annals of emergency medicine. 2013 Dec:62(6):584-91. doi: 10.1016/j.annemergmed.2013.04.010. Epub 2013 Jun 12     [PubMed PMID: 23769645]

Level 1 (high-level) evidence

[15]

Guirgis FW, Williams DJ, Kalynych CJ, Hardy ME, Jones AE, Dodani S, Wears RL. End-tidal carbon dioxide as a goal of early sepsis therapy. The American journal of emergency medicine. 2014 Nov:32(11):1351-6. doi: 10.1016/j.ajem.2014.08.036. Epub 2014 Aug 21     [PubMed PMID: 25205615]